Routine maintenance – Labconco Protector Multi-Hazard Glove Box 5065502 User Manual
Page 14
ROUTINE MAINTENANCE
15
Gloves
At least every three months (more often if subjected to heavy use), inspect the gloves for
signs of damage or wear. For replacement, refer to page 11.
Transfer Chamber Inner and Outer Door
Periodically check the closing tension of the door latches. Closing the latch handle should
require some firm force, indicating compression of the door seals. Seals, compression, and
latching force can be adjusted as follows:
1. Close and latch the door.
2. Using a 1/2" open-end wrench, turn the door adjustment screw (located between the door
and the latch bar) clockwise to increase seal compression and latching force. Turn screw
counterclockwise to decrease compression. Refer to figure 5.
Figure 5