Many excellent reference texts and booklets are currently available. The
following is a brief listing:
Laboratory Ventilation Standards
Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices.”
• Fume hoods should have a continuous monitoring device
• Face velocities should be between 60-100 linear feet per minute (lfpm)
• Average 2.5 linear feet of hood space per person
Occupational Health and Safety
U.S. Department of Labor
200 Constitution Avenue N.W.
Washington, DC 20210
(202) 523-1452
Industrial Ventilation-ACGIH
• Fume hood face velocities between 60-100 lfpm
• Maximum of 125 lfpm for radioisotope hoods
• Duct velocities of 1000-2000 fpm for vapors, gasses and smoke
• Stack discharge height 1.3-2.0 x building height
• Well designed fume hood containment loss, <0.10 ppm