Electrical requirements, Service line requirements, Space requirements – Labconco Protector XLE Fume Hoods 9970801 User Manual
Page 11: Your next step, Chapter 2: prerequisites
Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Self-Closing 60% Sash Opening 18"
Hood
Size
Standard Model Description
Face Velocity
Exhaust
Volume
Static
Pressure Loss
4'
48" Protector XLE Hood
60 fpm
310 CFM
0.03" H
2
O
80 fpm
420 CFM
0.06" H
2
O
100 fpm
520 CFM
0.09" H
2
O
5'
60" Protector XLE Hood
60 fpm
410 CFM
0.04" H
2
O
80 fpm
550 CFM
0.08" H
2
O
100 fpm
690 CFM
0.12" H
2
O
6'
72" Protector XLE Hood
60 fpm
510 CFM
0.06" H
2
O
80 fpm
680 CFM
0.10" H
2
O
100 fpm
850 CFM
0.15" H
2
O
7'
84" Protector XLE Hood
60 fpm
610 CFM
0.03" H
2
O
80 fpm
820 CFM
0.06" H
2
O
100 fpm
1020 CFM
0.09" H
2
O
8’
96" Protector XLE Hood
60 fpm
710 CFM
0.04" H
2
O
80 fpm
940 CFM
0.06" H
2
O
100 fpm
1180 CFM
0.10" H
2
O
Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a blower system.
Electrical Requirements
The Protector Hood models feature internal wiring for the fluorescent light
assembly and light switch. All internal wiring is terminated at the single point
wiring junction box for hook-up by a qualified electrician. The blower switch,
and light switch wires are also terminated at the single point wiring junction box
for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C: Protector XLE Specifications for the wiring diagram for proper
electrical installation.
Service Line Requirements
All service lines to the laboratory fume hood should be a minimum of 1/4"
outside diameter (3/8" diameter for a cold water gooseneck), copper (brass for
natural gas), and equipped with an easily accessible shut-off valve, should
disconnection be required. If the service line pressure exceeds 40 PSI, it must be
equipped with a pressure regulator to reduce the line pressure. Please check with
local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in Appendix B: Protector XLE
Dimensions.
Your Next Step
After you have determined that the location you have selected accommodates the
installation and operational requirements of your fume hood, you are ready to
begin installation. Proceed to Chapter 3: Getting Started.