Chapter 6: Maintaining Your Protector Demonstration Hood
Product Service 1-800-522-7658
35
Note: Enter Button stores information and +/- Buttons allow for scrolling.
1. Push the “ENTER” Button on the face of the alarm until the “SET
UP” Menu is displayed.
2. Scroll to “SET UP” and hit “ENTER”.
3. The PASSWORD MENU displays (The Password is 0000). Press
the ENTER button repeatedly until the CAL CONFIG MENU is
displayed.
4. In the CAL CONFIG MENU set the following:
CALIBRATION - CONFIGURATION MENU
SETTINGS
DISPLAY UNITS
FPM
* 48 FPM (60 fpm Operating inflow)
* 64 FPM (80 fpm Operating inflow)
LOW AIR ALARM “RED LED”
* 80 FPM (100 fpm Operating inflow)
LOW AIR CUTOFF
OFF
* 51 FPM (60 fpm Operating inflow)
* 67 FPM (80 fpm Operating inflow)
WARNING AIR ALARM “YELLOW LED”
* 83 FPM (100 fpm Operating inflow)
WARNING AIR RESET
3 FPM
HIGH AIR ALARM
OFF
LOWER AIR SAMPLE FLUCTUATIONS
OFF
HIGHER AIR FLUCTUATIONS
3%
LOWER / HIGHER AIR SAMPLE DIFFERENCE 3%
WARN TO ALARM AIR TIME
10 SECONDS
ALARM TO WARN AIR TIME
3 SECONDS
SHOW AIR FLOW
ON
SHOW TIME LINE OFF = DISPLAYS BAR
GRAPH
OFF
AUDIBLE ALARM
ENABLED
SENSOR DIFFERENCE
2%
SENSITIVITY 80%
* (Suggested Air Alarm Settings) Refer to the chart on the previous pages for
a range of air velocity settings that may be used.
5. To complete the CAL CONFIG, be sure to enter “DONE”. You
are returned to the Main Menu.