Notice – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual
Page 9

• Check the power source and the pump’s rating plate
to be sure that the power source and the equipment
match in voltage, phase, and frequency.
• Use only oil of the recommended type. Other oils or op-
erating fluids may cause damage of the pump or dan-
ger.
Use
special oils (see “Oil change”, pg. 32) for the
rotary vane pump, if, e.g.,
- acid vapors are pumped.
- extremely high aging stability is required.
- oxygen or other strong oxidizing agents are pumped.
• Take adequate precautions when handling pump fluids,
lubricants, and solvents. Use appropriate protective
clothing, safety goggles and protective gloves to avoid
excessive contact with the skin and possible skin irrita-
tions (including dermatitis). Do not inhale or swallow!
Ensure that the pump location is well ventilated and
that possible toxic effects of certain vapors are avoided.
Maintain adequate levels of hygiene and cleanliness.
• Comply with all relevant statutory requirements and
regulations concerning the handling, storage and dis-
posal of oil.
Note: Do not allow oils to enter the drainage system
or other bodies of water. Spillage can cause accidents
(slip hazard)!
Keep a minimum distance of 8 in (20 cm) between the
cooling fan and surrounding items (e.g., housing, walls,
etc.). Check fan regularly for dust/dirt. Clean fan guard if
necessary to avoid a reduction of ventilation.
The On/Off switch is located at the side of the terminal
box.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s connections (to avoid overpressure at the outlet,
NOTICE