Labconco Rapid Digestors 23080 User Manual
Page 13

TROUBLESHOOTING
14
Symptom
Probable Cause
Corrective Measure
Main pilot light does not
illuminate.
ON-OFF switch in off position
Turn switch to ON position
Main power fuse blown or
circuit breaker tripped
Replace fuse or reset circuit
breaker
25 Amp fuse on control unit
blown (8 Amp on Rapid
Digestor 4-Place)
Check continuity of fuse with
ohmmeter. Replace fuse if
blown. If fuse continues to
blow, locate cause.
ON-OFF switch defective
Replace switch. Refer to
Service Section for
replacement procedure.
No heat applied to unit.
Main pilot light
illuminated
Sensor defective
Check continuity of sensor
by placing ohmmeter across
pins 1 & 2 of female sensor
receptacle. If resistance
measurements show sensor to
be open, replace sensor by
following Service Section on
Sensor Replacement.
Indicator lamp not
illuminated.
Bulb defective
Check continuity of bulb
with ohmmeter and replace if
defective following Service
Section-Indicator Lamp
Replacement.
Defective resistor
Check continuity of resistor
for the row of heaters behind
indicator lamp not
illuminated using ohmmeter.
Defective ceramic heater
Check continuity of each
heater contained in the row
behind indicator lamp that is
not illuminated. Replace
defective heater following
Ceramic Heater Replacement
section.
Uneven heat applied to
unit (not applicable to
Rapid Digestor 4-Place
Strip heater defective
Check continuity of each
strip heater and replace
defective heater following
Strip Heater Replacement
section.