Detailed function of Components
1. Manifold Assembly. The hood features a molded, 6" diameter exhaust connection sized to
allow for a minimum of static pressure loss through the hood structure and one that still can
provide a good transport velocity through the exhaust system.
2. Condenser Head System. These cone shaped heads located above each heater provide an
essential part of the digestion process. The vapors from the boiling digestion medium are
condensed on the condenser heads and returned to the medium continually washing the sides
of the beakers in the process.
3. Heater System. Each 350 watt adjustable spring loaded heater provides an even distribution
of heat through the use of coiled heater wires embedded in a ceramic material.
4. Variable Heat Control Switch. Provides variable input from 20-100% of capacity to each
heater.
5. Elevation Control Knob. Adjusts beakers positions.
6. Main Power Switch. Turn apparatus on and off without need of changing variable heat
control switches.