Location Requirements
The Purifier Clean Bench should be located away from traffic
patterns, doors, fans, ventilation registers, fume hoods and any
other air-handling device that could disrupt its airflow patterns.
All windows in the room should be closed. Figure 2-1 shows the
optimum location for the bench.
L’hotte à flux laminaire Purifier doit etre située à l’écart de la
circulation, des portes, des ventilateurs, des registres de ventilation,
des sorbonnes et n’importe quel dispositif de traitement d’air qui
peut interrompre le circuit du courant d’air. Toutes les fênetres de
la chambre doit etre fermée. La figure 2-1 montre l’emplacement
optimal pour l’hotte.