Width (feet)
3
4
5
6
8
HEPA Filter Dims. (in.)
36 x 30 x 3
48 x 30 x 3
60 x 30 x 3
72 x 30 x 3
48 x 30 x 3
Avg. Velocity Reading Grid (inches)
12 x 12
12 x 12
12 x 12
12 x 12
12 x 12
Grid Distance from Diffuser (inches)
6
6
6
6
6
Grid Distance from Sidewalls (inches)
6
6
6
6
6
Average Velocity (FPM)
90 ±10
90 ±10
90 ±10
90 ±10
90 ±10
Work Area (ft
2
)
7.5
10
12.5
15
20
Exhaust Volume Range (CFM)
600-750
800-1000
1025-1225
1200-1500
1600-2000
Number of Laskin Nozzles needed
2
2
2
2
2
APPROX. SETTINGS FOR NEW FILTERS
Mag Gauge Reading
0.7-1.0
0.7-1.0
0.7 – 1.0
0.7 – 1.0
0.7 – 1.0
Motor HP
1-1/2
1-1/2
2-1/3
2-1/3
2-1/2
Speed Winding used
High
High
High
High
High
Motor Capacitator
15
15
15
15
15
Ultraviolet Light(s) OPTIONAL
G30T8
G30T8
G30T8
G64T5
G64T5
Fluorescent Light(s)
F25T8/
TL741
F32T8/
TL741
F32T8/
TL741
F40T8/
TL741
F32T8/