Model
34308 34300 34408 34400 34508 34500 34608 34600
Cabinet
Size
(in
feet) 3 3 4 4 5 5 6 6
Sash
Opening
(inches) 8 10 8 10 8 10 8 10
Starting Serial #
1
0705_
0705_
0705_
0705_
0705_
0705_
0705_
0705_
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5 55+/-5
55+/-5 55+/-5
55+/-5
Nominal
Average
Inflow
(FPM) 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5
Supply HEPA Data
Labconco
P/N
3838400 3838400 3838401 3838401 3838402 3838402 3838403 3838403
Exhaust HEPA Data
Labconco
P/N
3838500 3838500 3838501 3838501 3838502 3838502 3838503 3838503
Motor/Blower Data
Labconco P/N
2
3832200
3832200
3832201
3832201
3832202
3832202
3832203
3832203
Motor HP
1/2
1/2
1/2
1/2
3/4
3/4
3/4
3/4
Blower Size (inches)
10 x 8
10 x 8
10 x 8
10 x 8
10 x 8
10 x 8
10 x 8
10 x 8
Fluorescent/UV Lamp Data
Fluorescent Lamps (2 each)
F25T8
TL741
F25T8
TL741
F32T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
F40T8
SP41
F40T8
SP41
Light intensity (ft-candles)
3
90-100 90-100 90-100 90-100 90-100 90-100 90-100 90-100
UV Lamp
G30T8
G30T8
G30T8
G30T8
G30T8
G30T8
G30T8
G30T8
UV energy(uW-cm
2
)
4
240
240
240
240
240
240
240
240
Prefilter Data
Labconco
P/N
3858900 3858900 3858901 3858901 3858902 3858902 3858903 3858903
Postfilter Data
Labconco P/N
NA
NA
3837201 3837201
3837202
3837202
3837203
3837203
1.
The primary serial tag is on the lower outside edge of the right corner post.
The secondary serial tag is located on the front of the electronics module on
the top right side of the cabinet. The first two digits of the serial number are
the year of production, the next two are the month. The next 5 digits are the
sequence of production, and the letter following the serial number is the
revision level of the cabinet.
2.
Each motor must be programmed by Labconco for the appropriate width
cabinet.
3.
As per ANSI/NSF Std. 49 with a background of 15 ft-candles.
4.
Measured in the geometric center of the work surface.
5.
If the optional Canopy Alarm is installed