Working in the biosafety cabinet – Labconco Biological Safety Cabinets 34608 Series User Manual
Page 45

Chapter 5: Using The Cabinet
Product Service 1-800-522-7658
40
Testing the battery backup alarm/Replacing the battery:
1. Turn the Logic off.
2. Unplug the Logic from the electrical supply. The Red alarm LED
should light, and the audio alarm should sound. If this does not occur,
replace the 9 volt backup battery inside of the alarm module by
removing the single Phillips screw on the front of the module to
access the battery inside. Reinstall the alarm module.
3. Plug the Logic back into its electrical supply.
4. The audio and LED alarm will cease.
Working In the Biosafety Cabinet
Planning
• Thoroughly understand procedures and equipment required before
beginning work.
• Arrange for minimal disruptions, such as room traffic or entry into the
room while the cabinet is in use.
Start-up
• Turn off UV light if included.
• Slowly raise the sash until the bottom of the sash aligns with the sash
indicator decal located on the left side of the work area.
• Turn on the fluorescent light and cabinet blower if the Smart-Start
features have not been activated.
• Check the air grilles for obstructions, and note the filter gauge
reading.
• Allow the cabinet to operate unobstructed for 5 minutes.
• Wash hands and arms thoroughly with germicidal soap.
• Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber
gloves. Use protective eyewear. Wear a protective mask if
appropriate.
Wipe-Down
• Raise the sash to its full open position (approximately 21.75 inches or
552 mm). Mute the alarm by depressing the "Alarm Silence" switch.
• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.