Chapter 7: Troubleshooting
Product Service 1-800-522-7658
56
PROBLEM
CAUSE
CORRECTIVE ACTION
Airflow Alert goes
off and/or there is a
slight decrease in
filter life remaining
gauge
HEPA filter loading
The gauge reading steadily decreases as
the Cabinet is used.
Blockage of the
return air slots or
grille
Check all return air slots and grilles to
ensure that they are not blocked or
restricted.
Blockage of the
exhaust outlet
Ensure that the exhaust outlet is not
blocked or restricted.
Blockage
or
restriction under the
work surface
Ensure that the towel catch and plenum
beneath the work surface are
unobstructed.
Contamination of
work in the cabinet
Improper technique
or procedure for the
Biosafety Cabinet
See “Use of the Cabinet” section in the
manual.
Restriction of the
return air slots or
grille – blockage of
the exhaust outlet
Ensure that all return air slots, grilles
and the exhaust outlet are unobstructed.
External factors are
disrupting the cabinet
airflow patterns or
acting as a source of
contamination
See “Installation” section of this
manual.
Cabinet is out of
adjustment/HEPA
filter(s) are defective
Have cabinet recertified.