Appendix C
12
**Note**
The elevation correction chart supports new style Meth-
ane
ODORATORS
only. Instruments manufactured
before July 1997, that have not been upgraded or cali-
brated to a methane reference other than 1.00% will not
use the
elevation
correction charts. Consult the factory
for using older style
ODORATORS
at
elevation
. To
identify an
ODORATOR’S
style or to tell if an upgrade
has been performed, check the top panel to see if there
is a snap-in bezel or a clear window in the overlay cov-
ering the display (“LCD”). Older style
ODORATORS
have a snap-in bezel over the LCD.
ODORATORS
not
calibrated for 1.00% methane have correction charts
for that gas or reference level and can be identified by
the chart as the reading and actual values will be equal
at the reference level.