Typical elution times – Heath Consultants EI-5 Ethane Gas Identifier User Manual
Page 15

EI-5 Manual Rev A 12
Carrier gas regulator set at 6 P.S.I.
Gas
Approximate Time
Methane
10 - 15 Seconds
Ethane
30 - 70 Seconds
Propane
3 - 6 Minutes
Note: The times shown in the above table is the elapsed times between
switching the “Valve” to the “Time” position and the corresponding upscale
indications on the display.
Note: The material used in the separation column has the property of
absorbing and retaining water vapor. Thus the useful life of the separation
column is dependent on the amount of moisture present in the gases being
analyzed.
There may be a need to change the separation column after multiple uses.
The instrument should be returned to a qualified service center for column
replacement.
TYPICAL ELUTION TIMES