ERRORS AND WARNINGS
The instrument shows clear messages when erroneous condition appears. Messages are also
displayed when the obtained values are outside expected range. The beeper is playing a beep on
errors.
a) on zero reading
Light High:
There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low:
There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
No Light:
The instrument cannot adjust the light level. Please
check that the sample does not contain any debris.
Inverted cuvettes:
The sample and the zero cuvette are inverted.
Zero:
A zero reading was not taken. Follow the instructions of
the measurement procedure for zeroing the meter.
b) on sample reading
Under Range:
A blinking “0.0” indicates that the sample
absorbs less light than the zero reference. Check the procedure
and make sure you use the same cuvette for reference (zero) and
measurement.
- HI 993302 (2 pages)
- HI 991405 (2 pages)
- HI 991401-01 (2 pages)
- HI 983308 (2 pages)
- HI 983302N (2 pages)
- HI 982401 (4 pages)
- HI 981406 (2 pages)
- HI 981405N (2 pages)
- HI 981402C (2 pages)
- HI 981402 (2 pages)
- HI 981401N (2 pages)
- HI 981400 (2 pages)
- HI 146-99 (2 pages)
- HI 146-01 (2 pages)
- HI 7636 (2 pages)
- HI 7634-00 (2 pages)
- HI 2000 Series (2 pages)
- HI 84102 (20 pages)
- HI 84100 (12 pages)
- HI 8410 (20 pages)
- HI 7874 (2 pages)
- BL Series Pumps (17 pages)
- BL 983329 (2 pages)
- BL 983327 (2 pages)
- BL 983324 (2 pages)
- BL 983322 (2 pages)
- BL 983321 (2 pages)
- BL 983320 (2 pages)
- BL 983319 (2 pages)
- BL 983318 (2 pages)
- BL 983317 (2 pages)
- BL 983315 (2 pages)
- BL 983314 (2 pages)
- BL 983313 (2 pages)
- BL 982411 (2 pages)
- BL 981411 (2 pages)
- BL 932700 (2 pages)
- BL 931700 (2 pages)
- HI 324N (15 pages)
- HI 200M (2 pages)
- HI 180 (2 pages)
- HI 97500 (2 pages)
- HI 8060 (80 pages)
- HI 98186 (68 pages)
- HI 93732N (2 pages)