Hanna Instruments DiST 7 User Manual
Dist, 7 & dist
w w w . h a n n a i n s t . c o m
WARRANTY
DiST
®
7 & DiST
®
8 are warranted for one year against defects in
workmanship and materials when used for their intended purpose and
maintained according to instructions. This warranty is limited to repair or
replacement free of charge. Damages due to accident, misuse, tampering
or lack of prescribed maintenance are not covered. If service is required,
contact the dealer from whom you purchased the instrument. If under
warranty, report the model number, date of purchase, serial number and
the nature of the failure. If the repair is not covered by the warranty, you
will be notified of the charges incurred. If the instrument is to be returned
to Hanna Instruments, first obtain a Returned Goods Authorization Num-
ber from the Customer Service department and then send it with shipment
costs prepaid. When shipping any instrument, make sure it is properly
packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
Instruction Manual
DiST
®
7 & DiST
®
8
EC/TDS & Temperature meters
Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using the meter.
If you need additional technical information, do not hesitate to e-mail us
at [email protected].
These instruments are in compliance with the CE directives.
Remove the instrument from the packing material and examine it care-
fully. If any damage has occurred during shipment, immediately notify
your Dealer or the nearest Hanna Customer Service Center.
Each meter is supplied complete with:
• Calibration solution, 20 mL sachet
HI 70031 (1413 µS/cm) for DiST
®
7
HI 70030 (12.88 mS/cm) for DiST
®
8
• 4 x 1.5V AA batteries
• Instruction manual
Note: Conserve all packing material until the instrument has been ob-
served to function correctly. Any defective item must be returned
in its original packing.
PRELIMINARY EXAMINATION
DiST
®
7 and DiST
®
8 are portable EC/TDS and temperature
meters specially designed for analytical field applications.
These meters provide fast, accurate, on-the-spot measure-
ments, with a unique stability indicator right on the LCD.
All EC/TDS readings are automatically temperature compen-
sated (ATC), and temperature values can be displayed in °C
or °F units.
The EC/TDS conversion factor (CONV) is selectable by the user,
as well as the temperature compensation coefficient b (BETA).
The meters can be easily calibrated at one point.
A low battery symbol warns the user when the batteries are
to be replaced. In addition, the Battery Error Prevention
System (BEPS) automatically turns the meter off for avoiding
erroneous readings caused by low voltage level.
The built-in temperature and conductivity sensors are pro-
tected inside the cell cup.
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
1. Cell cup
2. SET/HOLD button
3. ON/OFF/MODE button
4. Low battery indicator
5. Stability indicator
SPECIFICATIONS
Range
0.0 to 60.0°C / 32.0 to 140.0°F
DiST 7
0 to 3999 µS/cm / 0 to 2000 ppm
DiST 8 0.00 to 20.00 mS/cm / 0.00 to 10.00 ppt
Resolution
0.1°C / 0.1°F
DiST 7
1 µS/cm / 1 ppm
DiST 8
0.01 mS/cm / 0.01 ppt
Accuracy
(@20°C/68°F)
±0.5°C / ±1°F
±2% f.s. (EC/TDS)
Typical
EMC Deviation
±0.5°C / ±1°F
±2% f.s. (EC/TDS)
Temperature
Automatic,
Compensation with
β
adjustable from 0.0 to 2.4%/ºC
TDS Factor
Adjustable from 0.45 to 1.00 (CONV)
Calibration
Automatic, 1 point
Calibration Solutions
DiST 7
HI 7031 (1413 µS/cm)
HI 7032 (1382 ppm with CONV=0.5)
HI 70442 (1500 ppm with CONV=0.7)
DiST 8
HI 7030 (12.88 mS/cm)
HI 70038 (6.44 ppt with CONV=0.5
or 9.02 ppt with CONV=0.7)
Environment
0 to 50°C (32 to 122°F); RH 100%
Battery
Type
4 x 1.5V AA
Battery Life
Approx. 2000 hours of use
Auto-off
After 8 minutes of non-use
Dimensions
120 x 53 x 81 mm (4.7 x 2.1 x 3.2”)
Weight
205 g (7.2 oz)
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.