beautypg.com

Operational guide, Functional description & specifications of hi 8734 – Hanna Instruments HI 8733 User Manual

Page 4

background image

7

6

• Each meter is supplied complete with a 9V battery. Remove the

battery compartment cover on the back of the meter (see
page 16). Install the battery while paying attention to its
polarity.

• Connect the probe to the meter securely

by aligning the pins with the socket
and pushing the plug in.

• Make sure that the meter has been calibrated before taking any

measurements (see "Calibration" section).

• Immerse the conductivity probe into the

sample, with the holes on the shaft
completely submerged.
If possible, use plastic beakers or con-
tainers to minimize any EMC interference.

• Tap the probe lightly on the bottom of the beaker to remove any

air bubbles which may be trapped inside the PVC sleeve.

• Turn the instrument on by pressing the ON/OFF key.

• For HI 8633 and HI 8734:

Take the temperature of the solution
with a ChecktempC or another accurate
thermometer.

Press the COND/TEMP key for HI 8633 or TDS/TEMP for HI 8734
to display the temperature and adjust the temperature knob to
that of the solution e.g. 20°C.

• For HI 8733:

Adjust the TEMPERATURE COEFFICIENT
knob to 2% to compensate for the
temperature effect of average solutions
(to determine exact value for a particu-
lar solution, see page 14).

OPERATIONAL GUIDE

°C

COND.

TEMP.

TDS

TEMP.

ºC

2.5

2

1.5

1

0.5

0

TEMPERATURE COEFFICIENT

1) Liquid Crystal Display
2) Measurement range selection keys
3) Manual temperature compensation knob
4) TDS/temperature selection key
5) ON/OFF key

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI 8734

Range

0.0 to 199.9 / 0 to 1999 mg/L

0.00 to 19.99 g/L

Resolution

0.1 / 1 mg/L

0.01 g/L

Accuracy

±1% Full Scale

(@ 20°C / 68°F)

excluding probe error

Typical EMC

±2 % Full Scale

Deviation
Calibration

Manual, 1 point, through trimmer

Temperature

Manual, 0 to 50°C (32 to 122°F)

Compensation

with

β

= 2%/°C

TDS factor

0.5

Probe (included)

HI 76301D with 1m (3.3') cable

Environment

0 to 50°C (32 to 122°F); RH max 100%

Battery Type

1 x 9V, alkaline (IEC 6LR61)

Life

Approx. 100 hours of continuous use

Dimensions

164 x 76 x 45 mm (6.5 x 3.0 x1.8")

Weight

250 g (8.8 oz.)