• Press the COND/TEMP key to display conduc-

tivity.

• The display should read the conductivity of the

calibrating solution at the reference tempera-
ture.
E.g. using HI7030/HI8030 with a so-
lution at 18°C, the reading will be
"12.88 mS".

PROCEDURE FOR HI8820:

• Pour a small quantity of the conductivity solu-

tion into a plastic beaker.

• Immerse the conductivity probe in the solution sub-

merging the holes of the sleeve (0.5 cm below).
Make sure that no air bubbles are trapped inside
the glass sleeve. Wait for 2 or 3 minutes for thermal
equilibrium.

• Switch the instrument on by pressing the ON/OFF

key.

• Select the appropriate conductivity range

e.g. "19.99 mS"

for HI7030/HI8030,

"99.9 mS"

for HI7034/HI8034,

"1999µS"

for HI7031/HI8031,

"199.9 µS"

for HI7033/HI8033.

• Set the temperature coefficient knob to

2%°C.

HI 7030

0

0.5

1 1.5

2

2.5

TEMPERATURE

COEFFICIENT