beautypg.com

Hanna Instruments HI 98300 User Manual

Page 2

background image

OPERATION
• Remove the protective cap, turn the

DiST

®

on, and immerse it into the

solution to be tested, without ex-
ceeding the maximum immersion
level.

• Stir gently and wait for the read-

ing to stabilize. DiST

®

automati-

cally compensates for temperature
variations.

• The reading on display is directly expressed in:

ppm (mg/L) for DiST

®

1 (HI 98300 / 301),

ppt (g/L) for DiST

®

2 (HI 98302), µS/cm for

DiST

®

3 (HI 98303) and mS/cm for DiST

®

4

(HI 98304).

CALIBRATION
• Immerse the DiST

®

in the proper calibration

solution (see Specifications table), without ex-
ceeding the maximum immersion level.

• Allow the reading to stabilize and with a

small screwdriver adjust the calibration trim-
mer until the display shows the required solu-
tion value.

For example HI 70032 is
the correct calibration so-
lution for HI 98301, and
the trimmer should be
turned until the display
shows "1382".

PROBE REPLACEMENT
For replacing the probe, contact your Dealer or
the nearest Hanna Service Center.
• Replace the built-in probe with a new:

HI 73301 for DiST

®

1

HI 73302 for DiST

®

2

HI 73303 for DiST

®

3

HI 73304 for DiST

®

4

BEPS AND BATTERY REPLACEMENT
DiS

®

T is provided with BEPS (Battery Error Pre-

venting System), which avoids any erroneous
reading due to low battery level, by switching the
meter off.
To replace the batteries, slide off the battery
compartment cover and replace all four 1.5V
batteries while paying attention to their polarity.

Batteries should only be replaced in a non-
hazardous area, and using the battery type
specified in this instruction manual.

ACCESSORIES

HI 70030P

12880 µS/cm calibration solu-
tion, 20 mL sachet, 25 pcs

HI 70031P

1413 µS/cm calibration solution,
20 mL sachet, 25 pcs

HI 70032P

1382 ppm (mg/L) calibration so-
lution, 20 mL sachet, 25 pcs

HI 70038P

6.44 ppt (g/L) calibration solu-
tion, 20 mL sachet, 25 pcs

HI 70039P

5000 µS/cm calibration solution,
20 mL sachet, 25 pcs

HI 70442P

1500 ppm calibration solution,
20 mL sachet, 25 pcs

HI 73301

Spare electrode for DiST

®

1

HI 73302

Spare electrode for DiST

®

2

HI 73303

Spare electrode for DiST

®

3

HI 73304

Spare electrode for DiST

®

4

HI 731326

Calibration screwdriver (20 pcs.)

DiST

is a registered Trademark of

Hanna Instruments

This manual is related to the following products: