For better accuracy, frequent calibration of the instru-

ment is recommended. In addition, the instrument

must be recalibrated whenever:

a) The conductivity probe is replaced.
b) When high accuracy is required.
c) At least once a month.

Calibration procedure
Immerse the tip of the probe in a beaker with HI 70031P

1413 µS/cm solution. The tip of the probe should be

about 2” deep in the liquid. The probe should be

kept at least 1” from the walls and the bottom of the

beaker. Adjust the trimmer to read 1.41 on the LCD.

CALIBRATION

BATTERY REPLACEMENT

PROBE MAINTENANCE

The HI 73331 probe is very easy to use. The only

precaution to be taken is to clean the area surrounding

the connector before removing the probe, in order to

avoid dirtying the connector.

Taking measurements
Connect the HI 73331 probe. The meter requires the

probe in order to do any measurement. Press the

ON/OFF button to turn the meter ON. Place the

probe to the desired depth and saturate the soil if

needed. The conductivity value, automatically com-

pensated for temperature, will be shown on the LCD.
The meter displays conductivity in mS/cm, which is

equivalent to dS/m.
Pressing the TEMP button, the temperature measured

by the probe will be shown while the button is held.
Note: Before taking any measurement make sure the

meter has been calibrated.

OPERATIONAL GUIDE

ACCESSORIES

HI 73331

Direct soil penetration conductivity probe

HI 7031M 1413 µS/cm (1.4 mS/cm) @25ºC cali-

bration solution, 230 mL bottle

HI 7031L

1413 µS/cm (1.4 mS/cm) @25ºC cali-

bration solution, 460 mL bottle

HI 731326 Calibration screwdriver, 20 pieces

To change the batteries, remove the 4 screws located on

the top of the meter.

Once the top has been removed, carefully replace the 4

batteries located in the compartment while paying atten-

tion to their polarity.

Replace the top, making sure that the gasket is properly

seated in place, and tighten the screws to ensure a

watertight seal.

TROUBLESHOOTING

• Batteries must be changed whenever the readings are

erratic, the LCD is dim, or the meter loses calibration.

• The meter cannot measure neither conductivity nor

temperature without the HI 73331 probe.

CALIBRA

TION

SPECIFICATIONS

:

y

ti

v

it

c

u

d

n

o

C

e

g

n

a

R

:

e

r

u

t

a

r

e

p

m

e

T

*

)

m

/

S

d

(

m

c

/

S

m

0

0

.

4

o

t

0

0

.

0

C

°

0

.

0

5

o

t

0

.

0

:

y

ti

v

it

c

u

d

n

o

C

n

o

it

u

l

o

s

e

R

:

e

r

u

t

a

r

e

p

m

e

T

m

c

/

S

m

1

0

.

0

C

°

1

.

0

:

y

ti

v

it

c

u

d

n

o

C

y

c

a

r

u

c

c

A

)

C

°

0

2

@

(

:

e

r

u

t

a

r

e

p

m

e

T

)

m

c

/

S

m

0

0

.

2

o

t

0

0

.

0

(

m

c

/

S

m

5

0

.

0

±

)

m

c

/

S

m

0

0

.

4

o

t

0

0

.

2

(

m

c

/

S

m

0

3

.

0

±

C

°

1

±

e

r

u

t

a

r

e

p

m

e

T

n

o

it

a

s

n

e

p

m

o

C

e

r

u

t

a

r

e

p

m

e

t

,

c

it

a

m

o

t

u

A

d

e

x

if

)

ß

(

t

n

e

i

c

if

f

e

o

c

C

°

/

%

2

@

n

o

it

a

r

b

il

a

C

l

a

u

n

a

m

t

n

i

o

p

1

e

p

y

T

y

r

e

tt

a

B

e

p

y

t

n

o

tt

u

b

V

5

.

1

x

4

e

b

o

r

P

)

d

e

d

u

l

c

n

i(

1

3

3

3

7

I

H

n

o

it

a

rt

e

n

e

p

)"

5

.

4

(

m

m

2

2

1

t

n

e

m

n

o

ri

v

n

E

C

°

0

.

0

5

o

t

0

.

0

s

n

o

is

n

e

m

i

D

)"

0

.

1

x

6

.

1

x

4

.

6

(

m

m

6

2

x

0

4

x

3

6

1

e

b

o

r

p

r

o

f

m

m

0

5

1

t

h

g

i

e

W

).

z

o

5

.

3

(

g

0

0

1

Recommendations for Users

Before using this product, make sure that it is entirely suit-

able for the environment in which it is used. Operation of this

instrument in residential areas could cause unacceptable

interferences to radio and TV equipment.

Avoid touching the probes at all times.

Any variation introduced by the user to the supplied equip-

ment may degrade the instrument’s EMC performance.

To avoid electrical shock, do not use this instrument when

voltages at the measurement surface exceed 24 VAC or 60

VDC. To avoid damages or burns, do not perform any mea-

surement in microwave ovens.

*

The meter gives indicative readings with lower accuracy be-

tween 4mS/cm and 10 mS/cm.