CHR xx
Change the instrument range according with the parameter
value (xx):
• xx=06 EC range
• xx=07 TDS range
• xx=08 NaCl range
The instrument will answer for these commands with:
> where: COMMANDS REQUIRING AN ANSWER The instrument will answer for these commands with: where the checksum is the bytes sum of the answer string sent as 2 ASCII Causes the instrument to send a complete set of readings in • EC and temperature reading on EC range. The answer string contains: • Meter mode (2 chars): • 06 - EC range • Meter status (2 chars of status byte): represents a 8 bit hexadecimal encoding. µ S, 1-mS) • 0x10 - temperature probe is connected
characters.
All the answer messages are with ASCII characters.
RAS
according with the current range:
• TDS, EC and temperature reading on TDS range.
• NaCl, EC and temperature reading on NaCl range.
• 07 - TDS range
• 08 - NaCl range
• 0x40 - TDS unit (0-ppm, 0-g/L)
• 0x20 - EC unit (0-
• 0x01 - new GLP data available
• 0x02 - new SETUP parameter
• 0x08 - AUTOHOLD is stable