Hanna Instruments HI 8043 User Manual

Page 4

background image

7

6

PROBE INITIAL PREPARATION

PROBE PREPARATION

All DO probes are shipped dry. To hydrate
the probe and prepare it for use, connect it to
the meter and proceed as follows.

1. Remove the red & black

plastic cap. This cap is
only used for protecting the
probe during shipment, and
can be thrown away.

2. Wet the sensor by soaking the bottom

(2.5 cm/1") of the probe in HI 7041S elec-
trolyte solution for 5 minutes.

3. Rinse the supllied membrane (HI 76407A)

with some electrolyte while shaking it gen-
tly. Refill with clean electrolyte.

4. Gently tap the sides of the

membrane with your finger
tip to ensure that no air
bubbles remain trapped in-
side. To avoid damaging
the membrane, do not tap
the membrane directly on
the bottom.

5. Make sure that the rubber

O-ring sits properly inside
the membrane cap.

6. With the sensor facing

down, screw the cap clock-
wise. Some electrolyte will
overflow.

When the probe is not used,
protect the membrane with the
supplied cap.

Shipping

cap

black

red

SPECIFICATIONS

Range

0.00 to 19.99 mg/L O

2

0.0 to 50.0°C (32 to 122 °F)

Resolution

0.01 mg/L O

2

0.1°C

Accuracy

±1.5% F.S. (O

2

)

±0.5°C

Typical EMC

±1.5% F.S. (O

2

)

Deviation

±1°C

Calibration

Manual, 2 point (zero and slope)

Temperature

Automatic, 0 to 30°C

Compensation

(32 to 86°F)

Probe

HI 76401 polarographic

with 3 m (10') cable (included)

Battery Type

1 x 9V

Life

approx. 100 hours of continuous use

Environment 0 to 50°C (32 to 122°F); RH max 95%

Dimensions

180 x 83 x 40 mm (7.1 x 3.3 x 1.6")

Weight

240 g (8.4 oz.)