Hanna Instruments reserves the right to modify the design,

construction and appearance of its products without advance

notice.

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

HI 764

09/4

Galvanic DO probe with built-in temperature sensor and

4 m cable

HI 764

09/10 Galvanic DO probe with built-in temperature sensor and

10 m cable

HI 76410-0 Protective sleeve for HI 76410 probes series
HI 76410A/P Membrane for HI 76410 probes series, 5 pcs.
HI 7042S

Electrolyte solution, 30 ml bottle

HI 710015 Shockproof rubber boot, blue color
HI 710016 Shockproof rubber boot, orange color
HI 731326 Calibration screwdriver (20 pcs)

PROBE MAINTENANCE & CLEANING

PROBE MAINTENANCE & CLEANING

PROBE MAINTENANCE & CLEANING

PROBE MAINTENANCE & CLEANING

PROBE MAINTENANCE & CLEANING

For a top performance probe, it is recommended to replace the membrane
every 2 months and the electrolyte once a month.
Proceed as follows:

• Unscrew the membrane by turning it counterclockwise.

• Rinse the supplied spare membrane (HI 764

09A/P) with some

electrolyte solution while shaking it gently. Refill with clean

electrolyte.

• Gently tap the membrane over a surface to ensure that no air

bubbles remain trapped.

• With the sensor facing down screw the cap clockwise to the end

of the threads. Some electrolyte will overflow.

If any deposit scales the sensors, gently brush the sensor surface with the
supplied scouring pad, while paying attention to not damage the plastic
body.