Hanna Instruments HI 84100 User Manual
Page 8
8
FUNCTIONAL DESCRIPTION
FRONT PANEL
FRONT PANEL
FRONT PANEL
FRONT PANEL
FRONT PANEL
REAR PANEL
REAR PANEL
REAR PANEL
REAR PANEL
REAR PANEL
INSTRUMENT DESCRIPTION
1) Titrant bottle
2) Liquid Crystal Display (LCD)
3) Keypad
4) Electrode holder
5) Peristaltic pump tube
17
• Place the probe holder on the top of the beaker and
secure by turning clockwise.
• Rinse the ORP electrode with deionized water, open
the refill cap and immerse into the stirring sample
until the teflon reference junction is completely
submerged. Be sure that the tip of the electrode is
not in the path of the stir bar.
• Purge the titrant into a waste beaker for 1-2 seconds
to ensure that the dosing tip is filled with fresh
titrant solution.
• Insert the dosing tip into the titrant tube sleeve. IT
IS CRITICAL THAT THE TIP BE IMMERSED
APPROXIMATELY 1.5 CM INTO THE SOLUTION BEING
TITRATED.
• Fill the 20 mL beaker up to the 5 mL mark with the
HI 84100-52 Acid Reagent, add the contents to
the 50 mL beaker containing the sample.
• Add the contents of one HI 84100-54 Stabilizer
powder packet to the beaker containing the sample.
• Press START STOP to start the titration. The LCD will
display “titr” as well as stirrer and pump tags to
indicate a titration in progress.
• At the end of the titration the sulphur dioxide
concentration is displayed in ppm (mg/L).