Table 3: examples of standard calibration gas, Sensor, Gas cylinder part number – ENMET EX-5185 User Manual
Page 12

EX-5185-PID
ENMET Corporation
10
Table 3: Examples of Standard Calibration Gas
Gas
Range
Alarm 1* Alarm 2* Alarm 3*
Sensor
Part Number
Span Gas
Isobutylene
Gas Cylinder
Part Number
Benzene
0 - 20 ppm
0.5 PPM
1.0 PPM
5.0 PPM
03028-010
10 PPM
03290-010
1, 3-Butadine
0 – 20 ppm
0.5 PPM
1.0 PPM
5.0 PPM
03028-010
10 PPM
03290-010
Carbon disulfide
0 – 20 ppm
1.0 PPM
5.0 PPM
10 PPM
03028-010
10 PPM
03290-010
Chlorobenzene
0 – 200 ppm
50 PPM
75 PPM
100 PPM
03028-011
100 PPM
03290-100
Cyclohexanone
0 – 200 ppm
25 PPM
50 PPM
75 PPM
03028-011
100 PPM
03290-100
Diethylamine
0 – 20 ppm
5.0 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Dimethyl acetamide
0 – 20 ppm
5.0 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Hydrazine
0 – 20 ppm
0.5 PPM
1.0 PPM
5.0 PPM
03028-010
10 PPM
03290-010
Isobutyl acetate
0 – 2000 ppm
100 PPM
250 PPM
500 PPM
03028-012
750 PPM
03290-750
Isopropyl acetate
0 – 2000 ppm
100 PPM
250 PPM
500 PPM
03028-012
750 PPM
03290-750
Isopropylamine
0 – 20 ppm
0.5 PPM
1.0 PPM
5.0 PPM
03028-010
10 PPM
03290-010
Methyl bromide
0 – 20 ppm
5.0 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Methyl ethyl Ketone
0 – 2000 ppm
100 PPM
250 PPM
500 PPM
03028-012
750 PPM
03290-750
Methyl mercaptan
0 – 20 ppm
0.2 PPM
0.5 PPM
1.0 PPM
03028-010
10 PPM
03290-010
Methylamine
0 – 20 ppm
5.0 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Naphthalene
0 – 20 ppm
5.0 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Octane
0 – 2000 ppm
100 PPM
250 PPM
500 PPM
03028-012
750 PPM
03290-750
Phenol
0 – 20 ppm
5 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Tetrachloroethylene
0 – 200 ppm
75 PPM
100 PPM
150 PPM
03028-011
100 PPM
03290-100
Toluene
0 – 2000 ppm
100 PPM
250 PPM
500 PPM
03028-012
750 PPM
03290-750
Trichloroethylene
0 – 20 ppm
5 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Trimethylbenzene
0 – 20 ppm
5 PPM
10 PPM
15 PPM
03028-010
10 PPM
03290-010
Vinyl; acetate
0 – 20 ppm
0.5 PPM
1.0 PPM
5.0 PPM
03028-010
10 PPM
03290-010
Vinyl chloride
0 – 20 ppm
0.2 PPM
0.5 PPM
1.0 PPM
03028-010
10 PPM
03290-010
*
N
OTE
:
The detectable range and the availability of span/calibration gas dictates that for some gasses the span/calibration gas will not
trigger the high alarm point.
*
N
OTE
:
EX-5185
alarm points are independent of a controller. When a controller is supplied by
ENMET
Corporation the alarm
points for the
EX-5185
and the controller have been set accordingly. See the manual of the controller the
EX-5185
sensor
transmitter has been connected to for proper adjustment of the alarm points of the controller with the
EX-5185
sensor
transmitter.