Figure 5: EX-5165 Maintenance Menu Flow chart
Apply Calibration Gas until signal value becomes
stable (about 1 to 4 minutes)
See
Figure 4
If cal is good display will indicate OK or Same
If cal is not within preset “range” display will
indicate Bad Sens
M
ENU
PV:
0
15.0
Cal OK
Same mV
Bad Sens
Span
S
ELECT
S
ELECT
OR
OR
N
OTE
: You can change the Calibration Gas Level.
H
OLD
the magnet over the S
ELECT
switch
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
See Section 5.2.2
S
ELECT
mA Span
M
ENU
S
ELECT
20.0
To change mA Span set point:
Tap the M
ENU
switch until mA Span is displayed
Tap the S
ELECT
switch to display the set point
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
To return to Normal Gas Display:
Tap M
ENU
switch until EXIT is displayed
Then tap S
ELECT
switch
To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point
The M
ENU
switch: changes digit indicated by underscore
cursor
The S
ELECT
switch: locks in the underscored digit and
moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay
Use M
ENU
and S
ELECT
switches as above to change time
delay. Between 0 and 5 seconds is allowed
If change is within range display moves to next menu
Λ
- Indicates increasing alarm
V - Indicates decreasing alarm
SetTDsec
0
Alarm3
S
ELECT
Λ 15.0
M
ENU
SetTDsec
0
Alarm1
M
ENU
S
ELECT
Λ 5.0
SetTDsec
0
Alarm2
M
ENU
S
ELECT
Λ 10.0