Figure 5: EX-5165 Maintenance Menu Flow chart

Apply Calibration Gas until signal value becomes
stable (about 1 to 4 minutes)

See

Figure 4

If cal is good display will indicate OK or Same

If cal is not within preset “range” display will

indicate Bad Sens

M

ENU

PV:

0

15.0

Cal OK

Same mV

Bad Sens

Span

S

ELECT

S

ELECT

OR

OR

N

OTE

: You can change the Calibration Gas Level.

H

OLD

the magnet over the S

ELECT

switch

The M

ENU

switch changes digit indicated by underscore cursor

The S

ELECT

switch locks underscored digit and moves to next digit

See Section 5.2.2

S

ELECT

mA Span

M

ENU

S

ELECT

20.0

To change mA Span set point:
Tap the M

ENU

switch until mA Span is displayed

Tap the S

ELECT

switch to display the set point

The M

ENU

switch changes digit indicated by underscore cursor

The S

ELECT

switch locks underscored digit and moves to next digit

To return to Normal Gas Display:
Tap M

ENU

switch until EXIT is displayed

Then tap S

ELECT

switch

To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point

The M

ENU

switch: changes digit indicated by underscore

cursor
The S

ELECT

switch: locks in the underscored digit and

moves to next digit

If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay

Use M

ENU

and S

ELECT

switches as above to change time

delay. Between 0 and 5 seconds is allowed

If change is within range display moves to next menu

Λ

- Indicates increasing alarm

V - Indicates decreasing alarm

SetTDsec

0

Alarm3

S

ELECT

Λ 15.0

M

ENU

SetTDsec

0

Alarm1

M

ENU

S

ELECT

Λ 5.0

SetTDsec

0

Alarm2

M

ENU

S

ELECT

Λ 10.0