beautypg.com

4 sensor replacement c, Figure 6: sensor replacement ma span – ENMET EX-5155 User Manual

Page 15

background image

EX-5155-MOS

ENMET Corporation

13


Figure 5: EX-5155-MOS Maintenance Menu Flow chart

5.3 Heater Voltage Settings

Heater Voltages are necessary for MOS sensors. They are preset at the factory and should not require field adjustment. Do not
adjust these voltages unless specifically instructed to do so by

ENMET

Corporation Technical Support Staff.

C

AUTION

:

Improper adjustment of heater voltages can damage sensors voiding any warranties and also alter the operating

characteristics of the sensor in such a way that the

EX-5155-MOS

may not respond to it’s target gas.

5.4 Sensor Replacement

C

AUTION

:

Area must be declassified during sensor replacement.

Sensors should be replaced when they can no longer be calibrated. Replacement sensor part numbers are listed in Section 6.0
of this manual. If you do not know the proper part number for your sensor, be sure to have the

EX-5155-MOS

serial number

available when contacting your Distributor or

ENMET

Corporation Technical Support.

To replace a sensor, it is not necessary to open the transmitter housing.
Remove the set screw from sensor housing base.
Unscrew the sensor housing cover and remove spacer.

Note the orientation of spacer.

Unplug the sensor form PC Board.
Plug new sensor into PC Board and replace spacer.

Replace spacer with grooved edge toward sensor housing cover.

Reassemble the sensor housing.
After the new sensor has been installed, it is suggested to allow the sensor to stabilize for 24 hours.

A Factory calibration must be performed.

After entering the Maintenance menu, advance to the Zero menu. Then while viewing the Zero menu, hold the magnet over the

M

ENU

switch for 2-4 seconds.

After 2-4 seconds, an F will appear on the far right hand side of the display. The F indicates that the instrument is in Factory
mode.

Perform the calibration Zero and Span procedures as outlined in Section 5.2. Be sure that the F is present when selecting the
Zero and Span functions.

The Factory calibration sets a calibration window for future standard instrument calibrations.

Only perform a factory calibration when installing a new sensor!!

Figure 6: Sensor Replacement

mA Span

M

ENU

S

ELECT

100

To change mA Span set point:
Tap the M

ENU

switch until mA Span is displayed

Tap the S

ELECT

switch to display the set point

The M

ENU

switch changes digit indicated by underscore cursor

The S

ELECT

switch locks underscored digit and moves to next digit

To return to Normal Gas Display:
Tap M

ENU

switch until EXIT is displayed

Then tap S

ELECT

switch

Set Screw

(0.050 inch, 1.27mm Hex Key)

Sensor Housing Base

Sensor

Sensor Housing Cover

P C Board

Spacer

Note the orientation of spacer for replacement

Grooved edge of spacer toward Sensor Housing Cover