Exit, Zero cal ok, Bad zero pv: 0 – ENMET EX-5120 User Manual

Page 12

background image

EX-5120

ENMET Corporation

10


5.2.2 Gas Span

It is recommended that the Zero Function be performed first.

Enter the maintenance menu. See

Figure 5,

EX-5120

Maintenance Menu flow chart.

1.

Tap the

M

ENU

switch once to show Span on the display.

2.

Tap the

S

ELECT

switch to perform a Span procedure. The display will alternate between the calibration gas concentration

and a signal level.

3.

Attach the associated calibration gas cylinder to the regulator and calibration cover. See to

Figure 4.

4.

Open the valve to apply the calibration gas to the sensor.

5.

Watch for the signal level to stabilize.

6.

Once the signal level has stabilized, the

EX-5120

will automatically lock in the calibration data and:

If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove calibration gas.
To exit maintenance menu tap the

M

ENU

switch until Exit appears, then tap the

S

ELECT

switch.

If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span.
Remove calibration gas. Tap the

M

ENU

switch until Exit appears, then tap the

S

ELECT

switch. Check span gas and

repeat calibration.

N

OTE

:

Some software revisions require the

S

ELECT

switch be tapped to accept the signal.

7.

Calibration is complete.

5.2.3 Exit Maintenance Menu

Exit maintenance, by tapping on the

M

ENU

switch until Exit appears on the display. Tap the

S

ELECT

switch to return to the

instrument Normal Gas Display.

Normal Display Mode

N

OTE

:

You can change the Calibration Gas Level. H

OLD

the magnet over the S

ELECT

switch for 2 – 4 seconds

The M

ENU

switch changes digit indicated by underscore cursor

The S

ELECT

switch locks underscored digit and moves to next digit

5%

M

ENU

H

OLD

the magnet over the M

ENU

switch for 2 – 4 seconds

to enter the Maintenance Menus

Exit

S

ELECT

Tap the S

ELECT

switch to return to the Normal

Display Mode.

See Section 5.2.3

M

ENU

Tap the magnet over the M

ENU

switch to cycle through Maintenance Menus

OR

Zero

Cal OK

If the Zero signal is within Preset Specs the EX-5120 will display

Cal OK,

See Section 5.2.1

If the Zero signal is not within Preset Specs the EX-5120 will

display Bad ZERO

N

OTE

:

Some software revisions require the

S

ELECT

switch be

tapped to accept the signal.

S

ELECT

M

ENU

Bad ZERO

PV:

0

S

ELECT

Tap the S

ELECT

switch to initiate Zero adjustment

Select

Menu