5.2.2 Gas Span
It is recommended that the Zero Function be performed first.
Enter the maintenance menu. See
Figure 5,
EX-5120
Maintenance Menu flow chart.
1.
Tap the
M
ENU
switch once to show Span on the display.
2.
Tap the
S
ELECT
switch to perform a Span procedure. The display will alternate between the calibration gas concentration
and a signal level.
3.
Attach the associated calibration gas cylinder to the regulator and calibration cover. See to
Figure 4.
4.
Open the valve to apply the calibration gas to the sensor.
5.
Watch for the signal level to stabilize.
6.
Once the signal level has stabilized, the
EX-5120
will automatically lock in the calibration data and:
If the Span is successful, Cal OK appears on the display momentarily, then advances to Alarm 1. Remove calibration gas.
To exit maintenance menu tap the
M
ENU
switch until Exit appears, then tap the
S
ELECT
switch.
If the sensor is outside of acceptable parameters, Bad Span is displayed momentarily, then returns to Span.
Remove calibration gas. Tap the
M
ENU
switch until Exit appears, then tap the
S
ELECT
switch. Check span gas and
repeat calibration.
N
OTE
:
Some software revisions require the
S
ELECT
switch be tapped to accept the signal.
7.
Calibration is complete.
5.2.3 Exit Maintenance Menu
Exit maintenance, by tapping on the
M
ENU
switch until Exit appears on the display. Tap the
S
ELECT
switch to return to the
instrument Normal Gas Display.
Normal Display Mode
N
OTE
:
You can change the Calibration Gas Level. H
OLD
the magnet over the S
ELECT
switch for 2 – 4 seconds
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
5%
M
ENU
H
OLD
the magnet over the M
ENU
switch for 2 – 4 seconds
to enter the Maintenance Menus
Exit
S
ELECT
Tap the S
ELECT
switch to return to the Normal
Display Mode.
See Section 5.2.3
M
ENU
Tap the magnet over the M
ENU
switch to cycle through Maintenance Menus
OR
Zero
Cal OK
If the Zero signal is within Preset Specs the EX-5120 will display
Cal OK,
See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5120 will
display Bad ZERO
N
OTE
:
Some software revisions require the
S
ELECT
switch be
tapped to accept the signal.
S
ELECT
M
ENU
Bad ZERO
PV:
0
S
ELECT
Tap the S
ELECT
switch to initiate Zero adjustment
Select
Menu