4.3 Sensor Hook-up

The MOS sensor is connected to the ISA-M control unit with three conductor wiring, use the correct oiltight fitting.
Two conductors supply heater current to the sensor. The third conductor is a signal wire. Size of heater wire depends
on the distance between the particular sensor and the control unit. See Table 4.

Sensor wires correspond to the normal wire code:

orange – heater
brown – heater ground
blue – signal

Figure 4: Internal View of Sensor Wiring

N

OTE

: The three color-coded wiring attachments must be performed by the user when replacing the sensor.

Table 4: Recommended Wire Gauge

Distance from Sensor to Control Unit

Recommended Wire Gauge

250 feet

16 AWG

350 feet

14 AWG

Longer Distances

Contact Factory

C

AUTION

: After you mount and install the ISA – M, you must adjust the sensor heater voltage (see section 4.4).

Sensor Wiring

F

OR

812 / 813 / 814 S

ENSORS

Position

Function

Wire Color

1

Signal

(Blue)

2

Heater

(Orange)

3

Ground

(Brown)

Optional

Splash Shield

Sensor Wiring

Terminal