2. measurement specifications – Elenco 3 1/2 Digit Cap./ Freq./ Trans. w/ Grey Boot User Manual

Page 6

background image

4-2. Measurement Specifications

Accuracy is +(% of reading + number in last digit) at 73

O

F (23

O

C)

+9

O

F (+5

O

C), <75% RH.

DC Voltage

400mV

100

m

V

+(0.5% of rdg+3dgt)

250V~

4V

1mV

+(0.5% of rdg+3dgt)

600V~

40V

10mV

+(0.5% of rdg+3dgt)

600V~

400V

100mV

+(0.5% of rdg+3dgt)

600V~

600V

1V

+(0.8% of rdg+2dgt)

600V~

Impedance: 10M

W

Note: Maximum voltage from “V

W

W

mA” socket to “COM” socket is 600V~, from “COM

socket to earth is 300V~.

AC Voltage

400mV

100

m

V

+(1.2% of rdg+3dgt)

250V~

4V

1mV

+(0.8% of rdg+3dgt)

600V~

40V

10mV

+(0.8% of rdg+3dgt)

600V~

400V

100mV

+(0.8% of rdg+3dgt)

600V~

600V

1V

+(1.2% of rdg+3dgt)

600V~

Frequency Response: 40-400Hz (40 - 100Hz on 600V range).

Resistance

400

W

0.1

W

+(1.0% of rdg+3dgt)

4k

W

1

W

+(1.0% of rdg+3dgt)

40k

W

10

W

+(1.0% of rdg+3dgt)

250V DC/ACrms

400k

W

100

W

+(1.0% of rdg+3dgt)

4M

W

1k

W

+(1.0% of rdg+3dgt)

40M

W

10k

W

+(2.0% of rdg+3dgt)

DC Current

400

m

A

0.1

m

A

+(1.2% of rdg+3dgt)

Protected by

4000

m

A

1

m

A

+(1.2% of rdg+3dgt)

Fast

40mA

10

m

A

+(1.5% of rdg+3dgt)

0.5A/250V

400mA

100

m

A

+(1.5% of rdg+3dgt)

Fuse

10A

1mA

+(2.0% of rdg+5dgt)

10A/250V Fuse

Caution: Maximum operation time is 15 seconds on 10A scale.

(5) Frequency and Duty Cycle Measurement

1. Connect the black test lead to the “COM” socket and the red test lead to

the “V

W

Hz” socket.

2. Set the selector switch to the “Hz” or “Duty” position.
3. Connect the probes across the source or load under measurement.

(6) h

FE

Test

1. Set the selector switch to the desired “h

FE

” position (PNP or NPN type

transistor).

2. Never apply an external voltage to the h

FE

sockets. Damage to the meter

may result.

3. Plug the transistor directly into the h

FE

socket. The socket is labeled E, B,

and C for Emitter, Base, and Collector.

4. Read the h

FE

(beta or DC current gain) in the display.

(7) Relative Value Display

Press the “REL” button on the meter to use the relative measurement
mode. The present value will be stored in memory. The new display value
is equal to the measurement value minus the stored value.

Example:

When you test the capacitance, you can use the Relative function to
eliminate the zero error.

(8) Auto Power Off and Disable

1. When the meter has been turned on for 15 minutes without any action

from the users, the meter will automatically turn itself off.

2. To disable the Auto Power Off function, press any button when the meter

is on or switch the selector switch.

Range

Resolution

Accuracy

Maximum Input

Range

Resolution

Accuracy

Maximum Input

Range

Resolution

Accuracy

Overload Protection

Range

Resolution

Accuracy

Overload Protection