Chromalox 8003 User Manual
Page 16

Chromalox 8003 User's Manual
13
Error Messages
The instrument has a self-diagnostic algorithm. When an
error is detected, the instrument shows “Err” indication
on the lower display while the upper display shows the
code of the detected error.
Error List
100
Write EEPROM error.
150
CPU error.
200
Write on protected memory.
201-2xx
Configuration parameter error. The two least
significant digit’s show the number of the
wrong parameter (ex. 209 Err shows an Error
on P9 parameter)
301
RTD input calibration error
305
TC input calibration error
307
RJ input calibration error
310
CT input calibration error
400
Control parameters error
500
Auto-zero error
502
RJ error
510
General error during calibration
procedure
Note:
1. When a configuration parameter error is
parameter only.
2. If an error 400 is detected, simultaneously push
the
▲ and ▼ pushbuttons for loading the
default parameters, then repeat control
parameter setting.
3. For all the other errors, contact your supplier.
8003 Manual
4/27/06, 2:24 PM
13