liquid level to detect
nonpolar liquids (e. g.
toluene)
liquid level to detect
polar liquids (e. g.
water)
Assembling the level sensor at the condensate flask 1000 ml
opening of the strap
➨
Attach both straps on the condensate flask.
➨
Position mounting plate under the straps. Assemble the liquid level sensor in a way that the sensor
becomes activated at the desired maximum liquid level.
➨
Close straps.
➨
Loosen the adjustment screw at the sensor.
➨
Adjust the sensor as close as possible to the condensate flask.
➨
Screw adjustment screw.
closing of the strap
mounting plate
straps
level sensor
Disassembling the level sensor from the mount-
ing plate
(e. g. to empty the condensate flask)
Unscrew adjustment screw at the level sensor and
remove sensor.
Reassemble the sensor in the original position.
When reassembling ensure that the sensor is posi-
tioned as close as possible to the condensate flask.
adjustment screw