VACUUBRAND MV 2 NT VARIO User Manual
Page 79
page 79 of 106
* If remote operation is selected or deselected, the user has to ensure
that no dangerous status of the system can occur due to the change of
the mode of operation, and must also take appropriate safety precau-
tions, especially if selecting remote operation interferes with a locally
operated active process.
** With command ”ECHO 1” a return value can be activated at write com-
mands. A return value is only given if the command is performed cor-
rectly.
*** After being switched on, the controller is in ”CVC 2” mode by default.
Send ”CVC 3” and ”STORE” to permanently set the controller RS 232C
commands to the extended set ”CVC 3000”.
Command
Operation
Parameter Description
OUT_SENSOR
1
2...9
internal sensor
external sensors (if connected)