Note: Flexible elements will shrink when evacuated.
• Connect hoses gas tight at inlet and outlet of the pump.
• Connect hoses gas tight at the pressure transducer.
• Ensure that no foreign objects can be drawn into the
pump.
• Check the power source and the pump’s rating plate
to be sure that the power source and the equipment
match in voltage, phase, and frequency.
Make sure ventilation is adequate to maintain recommend-
ed operating temperature. Keep a minimum distance of
2 in (5 cm) between the cooling fan and surrounding items
(e.g., housing, walls, etc.), or else install an external auto-
matic ventilation system. Check fan regularly for dust/dirt.
Clean fan guard grill if necessary to avoid a reduction of
ventilation.
MV 10 VARIO-B: Do not place the pump on soft surfaces
(e.g., rubber foam) during operation. This may reduce or
block the fans’ air supply.
Position the pressure transducer and its vacuum line in
such a way that condensate cannot flow towards the pres-
sure transducer.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s tubing (to avoid overpressure at the outlet, and
reduction of pumping speed at the inlet).
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Comply with all
applicable and relevant safety require-
ments (regulations and guidelines). Implement the re-
quired actions and adopt suitable safety measures.
NOTICE