Notice – VACUUBRAND MV 10 NT User Manual
Page 28
page 28 of 70
• Reduce the transmission of vibration. Pre-
vent mechanical load due to rigid pipelines.
Insert elastic hoses or flexible elements
as couplings between the pump and rigid
pipes.
Note: Flexible elements will compress or
flatten when evacuated if not designed for
use under vacuum.
• Hose connections at the pump inlet must al-
ways be gas tight.
Use a suitable adapter to connect a vacuum
hose (caoutchouc) 15 mm ID (see ”Accesso-
ries / Spare parts”, pg. 34), if necessary.
Use connecting hoses with large diameter and
keep them as short as possible to avoid flow
losses. Locate the pump as closely as possi-
ble to the application.
Secure hose connections at the pump ap-
propriately, e.g., with hose clamps, to protect
against accidental detachment.
Use a suitable valve (see “Accessories / Spare
parts”, pg. 34) to isolate the pump from the
vacuum application. This is to allow the pump
to warm up before pumping condensable va-
pors and to clean the pump after use before it
is switched off.
NOTICE
NOTICE
NOTICE
✔
✘
inlet