beautypg.com

VACUUBRAND RP 35C VARIO User Manual

Page 18

background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999019 / 22/12/2009

page 18 of 29

Cleaning and inspecting the pump heads:

RP 12C and RP 15C VARIO (with MD 4C)

Please read the whole chapter ”Replacing diaphragms and valves” before
starting maintenance.

Partially the pictures show pumps in other versions. This doesn´t influence replacing diaphragms and
valves of the pump.

Set of seals (diaphragms and valves) for MD 4C ..................................................................... 69 68 15

If necessary, disassemble diaphragm pump MD4C from pumping unit.

➨ Use open-ended wrench (w/f 17) to remove union nuts.

➨ Use open-ended wrench (w/f 15) to turn elbow fitting 1/4 of a

turn, remove hose.

☞ Do not remove the elbow fitting from the pump head. Through

reassembly a leak may result.

➨ To check valves use hex key to remove four socket head

screws from pump head and remove upper housing (hous-
ing cover with housing cover insert), head cover and valves.

☞ Never remove parts by using a spiky or sharp-edged tool

(e.g. screw driver), we recommend to use a rubber mallet or
compressed air (to be blown carefully into port).

➨ Remove head cover from housing cover insert and check

valves. Note position of valves and remove.

☞ Replace valves if necessary.

☞ Use petroleum ether or industrial solvent to remove depos-

its. Do not inhale.

This manual is related to the following products: