Safety during operation – VACUUBRAND PC 8 m. RC 6 User Manual
Page 17
![background image](/manuals/715226/17/background.png)
reactions of the pumped substances with each other
and with the pump’s materials.
Safety during operation
➨
Adopt suitable measures to prevent the release of dan-
gerous, toxic, explosive, corrosive, noxious or polluting
fluids, vapors and gases. To prevent any emission of
such substances from the pump outlet, install an ap-
propriate collecting and disposal system and take pro-
tective action for pump and environment.
➨
You must take suitable precautions to prevent any for-
mation of explosive mixtures in the oil reservoir (PC 8
with RC 6), in the pump chamber, or at the outlet of the
pump. In case, e.g., of a diaphragm failure, mechani-
cally generated sparks, hot surfaces or static electricity
may ignite these mixtures. Use inert gas for gas ballast
or venting, if necessary.
➨
Drain appropriately or otherwise remove any potential-
ly explosive mixtures at the outlet of the pump, or dilute
them with inert gas to non-explosive concentrations.
➨
Never operate this pump if it has a damaged cord or
plug.
+
If the pump is not working properly, has been dropped
or has fallen into water, contact your pump service pro-
vider.
+
Prevent any part of the human body from coming into
contact with vacuum.
+
Never aspirate liquids or dust into the pump.
+
Make sure that the exhaust pipeline cannot become
blocked.
+
Check the overpressure safety relief device at the ex-
haust waste vapor condenser at appropriate intervals.