Setting up and installing the equipment, Notice – VACUUBRAND PC 3001 basic pro User Manual
Page 12
page 12 of 54
+
Comply with all notes on correct vacuum and electri-
cal connections; see section „Use and operation“, pg.
+
Do not use the pump to generate pressure.
+
The pumps are designed for
ambient temperatures
during operation between +50°F and +104°F (+10°C
and +40°C). Periodically check maximum tempera-
tures if installing the pump in a cabinet or a housing.
Make sure ventilation is adequate to maintain recom-
mended operating temperature. Install an external au-
tomatic ventilation system if necessary. If pumping hot
process gases, make sure that the maximum permitted
gas inlet temperature is not exceeded. The maximum
permitted gas inlet temperature depends on several
parameters like inlet pressure and ambient tempera-
ture (see „Technical data“, pg. 23).
+
Particles and dust must not enter the pump.
+
Do not pump liquids.
• Ensure that the pump, and any accessories in the flow
path are chemically resistant to the pumped substanc-
es prior to operation.
Use the equipment
only as intended, that is, for genera-
tion and control of vacuum in vessels designed for that
purpose. Any other use will automatically invalidate all
warranty and liability claims. Remain aware of safety and
risks.
Setting up and installing the equipment
➨
Equipment must be connected only to a
suitable elec-
trical supply and a suitable ground point. As such,
the plug must be plugged into an outlet that is properly
grounded. Failure to connect the motor to ground may
result in deadly electrical shock.
NOTICE