The controller CVC 2000 can be adapted to the specific application by choosing the
appropriate mode, vacuum control with or without automatic, continuous pump-
ing with preset pumping speed or VACUU•LAN: The components of the chemistry
vacuum system (cooling water valve, air admittance valve) have to be preset once
only. The mode as from last operation and the preselected values (e. g. for pressure,
speed or time for switching off) are stored. In case of similar operation conditions it is
possible to start immediately, if the preselections are chosen appropriately.
How to configurate the controller
Vacuum control:
➨ Position control on preset pressure value.
➨ The pumping speed is adapted to the occurring gas/vapour,
so that the preset pressure value is kept constantly.
➨ Operating in equilibrium by continuous adaption of the con-
troller, no abrupt switching of valves.
Continuous pumping with preset pumping speed:
➨ Simple suction, suck liquids by siphon, drying, sucking off
excess, etc.
➨ Suction with regulated pumping speed.
➨ Pumping down until the ultimate vacuum of the pump is
achieved (e. g. to dry a system).
➨ System controlled by an interface: temporally variable pump-
ing speed to optimize the process conditions (e. g. to dry
sensitive substances).
➨ Einstellung ”HI”: Optimum pumping speed at every pressure.
Vacuum control in automatic mode:
➨ Searching the boiling point and adjusting the set point auto-
matically if the boiling point changes (due to temperature
change, change of ingredients of solutions).
➨ Adjusting the set point automatically after manual setting of
the boiling point (additional automatic mode in operation mode
vacuum control).
➨ Automatic switching off after complete evaporation of the
solvent if preselected.
In all modes:
☞ If cooling water and/or air admittance valve have been con-
nected and preselected:
➨ Process-adapted operation of the cooling water valve.
➨ Key-driven venting or adjustment of the set point by venting
if process control is active.
The vacuum controller CVC 2000 offers three modes depending on
-
the components of the VACUUBRAND chemistry vacuum system which are con-
nected to the system.
-