Triner Scale RAVAS-310 User Manual
Page 26

26
PARAMETERS:
Parameter
Function
Settings
Default
US
01
Start-up unit (and print units)
1=kg / 2=lb
2
02
Smallest graduation step for
multi-range
0.1/0.2/0.5…………10/20/50
0.5
03
Largest graduation step for
multi-range
0.1/0.2/0.5 ……….. 10/20/50
2
04
Number of graduations for
every range
0000-9900 divisions
1000
05
Weighing capacity system (full
scale)
0000-99999 units
5000
06
Motion tolerance for stable
0-32
1
off 0.5 grad./sec
1 grad./sec 2 grad./sec
4 grad./sec 8 grad./sec
16 grad./sec 32 grad./sec
07
Filter size
0-12
8
0=off
1=light filtering, 12=heavy filtering
08
Auto zero range
0=off 0.5 division
0.5
1=division 3 divisions
09
Zero range positive (+)
0-100% (approved 2%) of span
10
10
Zero range negative (-)
0-100% (approved 2%) of span
10
11
Test Function
BASIC ADC Counts
bASIC
10x Resolution
12-13
Not used
14
Start-up number to add in
sampling mode
1-2-5-10-20-50-95
10
15
Units switch mode active
Yes / No
Yes
16
Setpoint function
0-4
0
0=not used
1=overload gross (only 1 setpoint used)
2=overload gross (only 1 setpoint used)
3=Printer (without date/time/switched supply)
4=not used
17
Application
Basic (standard) or Peakhold (Phold)
bASIC
18
Gravity value working area
9.750-9.850
9.797
19
Key function
Remote - Local - Both
Local
20
Baudrate comport 1
600-1200-2400-4800-9600-19200
9600
21
Databits comport 1
7-8
8
22
Parity comport 1
none/odd/even
none
23
Stopbits comport 1
1-2
1
24
Not used