AC2500, AC25HD
60 seconds
AC2501, AC25HD, AC2504
60 seconds
AC5000, AC50HD
90 seconds
AC5001, AC50UV, AC5004
90 seconds
360401 with interface module
105 seconds
380401, 380402, 384UV5 with interface module 180 seconds
360801 with interface module
125 seconds
380801, 380802, 388UV5 with interface module 200 seconds
Note: Make sure the calorimeter delay is less than the time out setting you will make in section 10.
K. Press and hold the OFF/CANCEL button until the display blanks to save the settings.
L. Press the ON/SELECT button to turn the H410 back on.
6. To Select a Range:
Note: AUTO range is not available in the energy mode for calorimeters.
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the
display blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings
are saved they will be lost.
A. Press the RANGE button. The analog scale (if active) will disappear and the available ranges will
cycle through the display. The ranges available for calorimeters are in the following table:
Model
Astral 25mm
Astral 50mm
Large Aperture 100mm
with PN10735 Interface
Module*
Large Aperture 200mm
with PN10747 Interface
Module**
Mode
Power Energy Power Energy Power Energy Power Energy
Range
10mW
100mW
1W
10W
AUTO
10mJ
100mJ
1J
10J
300mW
3W
30W
AUTO
300mJ
3J
30J
300mW
3W
30W
AUTO
300mJ
3J
30J
300mW
3W
30W
AUTO
300mJ
3J
30J
* With a PN10748 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 50 W or
150J.
** With a PN10769 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 100 W or
300 J.
B. To select a range, press the ON/SELECT button when the range you desire appears in the display. The
selected range will be activated and the H410 will return to the last operational state.
C. Press and hold the OFF/CANCEL button until the display blanks to save the settings.
D. Press the ON/SELECT button to turn the H410 back on.
Note: When using PN10748 or PN10769, 10X attenuators, you may want to use an attenuation factor of
10 so the displayed reading on the S310 is correct. However the range designation may be
incorrect if the power or energy reading exceeds the upper limit of the range. See the Group
Settings section for details.
7. To Select a Measurement Mode:
The modes available for the calorimeters are: Energy (J) and Power (W).