LABEC AA9 User Manual
Page 2

Introduction
Congratulations on the choice of an Australian made quality product. Labec products are
manufactured, tested and calibrated to meet published standard specifications under our strict
quality assurance guidelines.
This Instruction Manual is for the guidance of operators of Labec Autoclaves and should be read
before the autoclave is connected to the electricity supply.
It is hoped that this manual will supply all the information that the customer should require for
satisfactory operation of this product. If, however, there are any questions that remain unanswered
please contact the Service Department via the website enquiry page and include the serial number
of the unit.
Unpacking
Remove all packing and protective wrapping from both interior and exterior of the autoclave. Check
the autoclave for any possible transit damage. Ensure all ordered accessories are present. If any
physical damage or shortages are evident, do not discard the packaging material until the autoclave
is inspected by the distributor, agent or manufacturer.
NOTE: All claims for shortage or damage must be made within fourteen days (14) from delivery.
Subject to our standard published conditions of sale, we have reasonable grounds to believe that we
have ensured, so far as is reasonably practical, that the products listed in our catalogue and
brochures have been designed and constructed so as to be safe and without risk to health when
properly installed and used in their environment by appropriate and trained personnel, and where
applicable, in accordance with our published instructions.
Installation
Electrical
This equipment must be tagged and tested according to AS/NZS3760:2010 prior to use and
thereafter every 6 months dependent upon the environment.
A qualified electrician should connect the power supply through the conduit at the back of the
autoclave. Access to the terminal block is obtained by removing the right hand side panel of the
autoclave. The frame of the autoclave must be sufficiently earthed using the earth terminal
located adjacent to the terminal block.
It is preferable to locate the autoclave close to a power point and recommended that double
adaptors are not used. Check the total wattage if connecting to multipoint outlets. Check the rating
plate for power requirements. Installation is to be carried out by a qualified electrician in accordance
with the power requirements of the product specifications.
See Appendix B attached.
Plumbing
Connect all plumbing in accordance with local authority’s regulations.
Location
Select a location free from draught and away from direct sunlight or other heat source. Proximity to
a drain for:
1.
Waste from the manually operated drain.
2.
The outlets servicing the pressure release valve and air vent
3.
Adjustable steam release valve at the bottom right hand rear of cabinet.
Laboratory Equipment Pty Ltd Instruction Manual Autoclave AA9 Bench Top Model