Т³гж 10, Control panel of, Control panel of 12l (fig 4-1) – LABEC AA-12BS User Manual
Page 9

Operation manual
Series Steam Sterilizers
5
4. Control Panel of
12L
Control Panel of 12L (Fig 4-1).
Fig 4-1
1
2
3
4
6
12
11
10
8
7
5
9
4
SELECT
key
5
PACKAGE
key
6
Test
key
1
Pressure Display
It displays chamber relative pressure in real time
during a cycle. The unit is per Bar.
2
Temperature
It displays chamber temperature during a cycle.
The unit is per Degrees Centigrade
3
State display/Error Codes
It displays all the states of autoclave. ( Look at the
state form) Error Codes when alarm goes on. You
may correct the error accordingly by looking it up
in the page of error codes and its corrections.
Caution: In case of faults, please contact your
distributor or the manufacturer immediately.
Program choose key. It is used for choosing the
temperature. You can choose the right program
according to the temperature that instrument can
+
afford. You can also choose B program, it can
sterilize medical cotton.
Package choose key. It is used for choosing two types mode of package. According to fact,
you can choose wrapped or unwrapped. Non-package use 3 times vacuumize, pakage use
B-standard vacuumize.
It is used for choosing the test program. If you push the button the indication light of B&D
goes on, and it means that you have chosen this testing program when you push again, the
indication light of Leak goes on, it means that you have chosen this program.