Operation manual
Series Steam Sterilizers
7
5. Operation of
SEA 12L
Please connect the power cord before you start the operation. Press the main power switch at
right lower corner of the front panel. The power switch light turns on indicating that power is
connected. It will indicate
Ld
when pass examination.
5.1 Fill water
If the water level is lower than the lowest water level.When the power switch is on, select the
sterilize program, the
water
indication light may glitting while alarm Beeps. At this time,
you need to refill the water reservoir.Please open the water tank and pour water into it by
container.Please pay attention to the water level during filling water,you should stop filling
when the surface is close to the red water level mark.
Fig 5-1
Caution:
1). Distilled water must be used to prolong life of the machine.
2). Don't tilt the autoclave when there is water in the water reservoir.
3). The condenser water collector should be emptied frequently. It is suggested
that the collector be emptied when refilling the water reservoir.
(Please refer to item 9.2 draining the water.)