beautypg.com

Cleaning & maintenance – Guardian Technologies AC4220 User Manual

Page 5

background image

GENERAL CLEANING

Always unplug the unit before cleaning or servicing in any way.
Caution: Do NOT use gasoline, benzine, thinner, harsh cleaners, etc. on and/or in the unit

while cleaning as they will damage the product. NEVER use alcohol or other solvents.
Note: Do NOT stick any foreign objects or your fingers inside the unit.
1.
Wipe the surface of the unit with a clean, dry soft cloth as needed.
2. Clean the Combination Filter monthly with the brush of your vacuum cleaner.

a. Remove Combination Filter from unit.
b. Remove the Pre-Filter/Charcoal layer

from the Velcro strips on the HEPA Filter

layer (Figure 2).

c. Clean both layers with the brush attachment

of your vacuum cleaner to remove lint and dust.

d. After cleaning, reattach the filter layers with

the Velcro strips and return filter to unit.

If the HEPA Filter layer turns from white to grey or black, it should be replaced with a new one. If

the HEPA Filter is still white, it does not need to be replaced.

FILTER REPLACEMENT

To maintain the performance of the air cleaner, it is recommended that the filter be replaced about

every 6 months. The replacement requirements will vary depending on usage and environmental

conditions. It’s time to replace the filter when

1. Filter Replacement Indicator light on the control panel will illuminate indicating that it is time

to replace your filter (Figure 3).

OR
2.
The HEPA Filter has changed from white to grey or black.

IMPORTANT:

Do NOT try to wash and re-use the Combination Filter. Please contact

Customer Service at

1.866.603.5900 or visit us at www.guardiantechnologies.com to

purchase a replacement filter.

Replacement Filter: Filter D (FLT4220)

1. Remove and replace filter according to “Directions For Use” on page E-6.
2. Reset Filter Replacement Indicator. Insert pen tip into HEPA reset area on control panel next to

UV-C power button to reset the indicator (Figure 4).

E-8

E-9

CLEANING & MAINTENANCE

4. The UV-C Bulb is removed by carefully grasping the bulb with a clean cloth or tissue and pulling

the bulb straight out of the unit (Figure 5).

5. The UV-C Bulb is reinstalled by sliding it, horizontally, back into the slots.

IMPORTANT: The UV-C Bulb should be replaced every 10-12 months

depending on use. Please contact Customer Service at

1.866.603.5900 or

visit us at

www.guardiantechnologies.com to purchase a

replacement bulb.
Replacement UV-C Bulb: LB4220

Pull UV-C Bulb

straight out of unit.

1. Lift and remove the

Front Cover from the unit.

2. After removing the

Front Cover from the

unit, remove the

Combination Filter.

3. After removing the

Combination Filter, use a

Philips head screwdriver to

remove two (2) screws and

detach the UV-C Shield.

Figure 5

Figure 2

CLEANING & MAINTENANCE

UV-C BULB REPLACEMENT

Figure 3

Filter Replacement

Indicator Light

HEPA

Reset

Figure 4