Calibration of the analogue output

1. Mounting the module

The SG module has to be installed in a cabinet in compliance with (inter)national electrical
regulations. Connection details are printed on the upper face of the amplifier housing.
The amplifier module supports 6 wire technique. If you are using a 4 wire load cell, the
“Exc +” and “Sen +” terminals must be shorted together as well as the “Exc –“ and “Sen –“
terminals.

2. Adjustment settings

All DIP-switches and potentiometers are positioned near the top edge of the housing.

3. Setup DIP-switches

Low pass filter:

By combining the six DIP-switches the low pass filter can be set from 0.25 Hz up to 32 Hz.
Factory default is 32 Hz.

DIP-switch ON

none

1

2

3

4

5

6

all

Cut-off frequency

(Hz)

32

16

8

4

2

1

0.5

0.25

Zero setting:

The six DIP-switches permit the off-set to be coarsely set in steps of 2.5% of 3.2 mV/V over the
range + or – 80%. The 25-turn potentiometer permits trimming to <0.002%.

DIP-switch ON

Pot.

none

1

2

3

4

5

all

Pol.

Relative Offset

(%)

0-3

0

2.5

5

10

20

40

80

+/-

Input at Zero V

out

(mV/V)

-0.1

0

0.08

0.15

0.3

0.6

1.2

2.4

+/-

Gain:

The five DIP-switches permit the relative gain factor to be set in steps of 2 over the range 1 to
64. The 25-turn potentiometer permits trimming to any gain factor <0.001x.

DIP-switch ON

Pot.

none

1

2

3

4

5

all

Rel. gain factor

(f)

0-2x

1x

2x

4x

8x

16x

32x

64x

Input at 10V V

out

(mV/V)

-

>3.3

2.7

1.8

1.1

0.6

0.3

0.17

Current output:

The I

out

range can be set to 0 – 20 mA or 4 – 20 mA.

FLINTEC

...the right weigh

Manual LAC 65.1

Page 3

5

10

20

40

80

160

320

0-3

2,5

5

10

20

40

-/+

1-2

2

4

8

16

32

0-20/4-20

mA

Rel.Gain(x)
SET SPAN

% of FS-in
SET ZERO

(ms)

L-P-FILTER

DIP-switches for

Low Pass Filter

0.5 - 32 Hz

DIP-switches and potentiometer

for zero setting

DIP-switches and potentiometer
for gain setting

DIP-switch for
Current Output
0 - 20 or 4 - 20 mA

OFF

ON