Premier series monitor, Calibration – Eagle Microsystems GD-1000 User Manual
Page 4

EAGLE MICROSYSTEMS. INC.
Instrumentation for Measurement and Control
4
copyright 2/02
ALARM SCREEN:
Press the pushbutton under CRT or DGR, to configure. Press ESC to return.
1
CRT
Critical alarm screen to set value for relay contact.
2
DGR
Danger alarm screen to set value for relay contact.
3
4
ESC
Go back to the monitor setup screen.
CRITICAL ALARM SETUP SCREEN:
From the critical alarm setup screen you can view the current value. To
change, press the pushbutton under UP or DWN to change the value. Press
the pushbutton under SET to enter the new value and return back to alarm
setup screen.
1
UP
To increase the value.
2
DWN
To decrease the value.
3
4
SET
Go back to the alarm setup screen.
DANGER ALARM SETUP SCREEN:
From the danger alarm setup screen you can view the current value.
To
change, press the pushbutton under UP or DWN to change the value. Press
the pushbutton under SET to enter the new value and return back to alarm
setup screen
1
UP
To increase the value.
2
DWN
To decrease the value.
3
4
SET
Go back to the alarm setup screen.
Premier Series Monitor
CALIBRATION:
SELECT CALIBRATION
From the MONITOR SETUP SCREEN, press the pushbutton under CAL.
1
CAL
Enter calibration screen..
2
ALM
Enter alarm select screen.
3
4
RUN
Go back to the main run screen
CALIBRATION SCREEN:
Press
the pushbutton under CH1 or IOUT
.
Press ESC to return.
1
CH1
Press to calibrate 4-20mA analog input..
2
3
IOUT
Press to calibrate 4-20mA analog output.
4
ESC
Will ask to set new password yes /no
SEL
SO2
CH1 ALM
CRT
DGR
ESC
CRITICAL
10.0
UP
DWN
SET
DANGER
2.0
UP
DWN
SET
MONITOR SETUP
CAL
ALM
RUN
CALIBRATION
CH1
IOUT
ESC