Buffalo Filter® VisiClear™ Operator’s Manual

Table 3

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

The model VisiClear™ is intended for use in the electromagnetic environment specified below.

The customer or user of the model VisiClear™ should assure that it is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance Level Electromagnetic Environment - Guidance

Portable and mobile RF communications

equipment should be used no closer to any

part of the Model EVL including cables,

than the Recommended separation distance

calculated from the equation applicable to

the frequency of the transmitter.

Radiated RF

3 V/m

d = 1.7 √P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5

GHz

3 V/m

d = 2.3 √P 800 MHz to 2.5 GHz

3 Vrms

d = [3.5/V1} √P

Conducted RF

IEC 61000-4-6

150 kHz to 80

MHz

Where P is the maximum output power

rating of the transmitter in watts (W)

according to the transmitter manufacturer

and d is the recommended separation

distance in meters (m).
Field strengths from fixed RF transmitters,

as determined by an electromagnetic site

survey, should be less than the compliance

level in each frequency range.
Interference may occur in the vicinity of

equipment marked with the following

symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structure, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site

survey should be considered. If the measured field strength in the location in which the Model EVL is used

exceeds the applicable RF compliance level above, the model VisiClear™ should be observed to verify normal

operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or

relocating the model VisiClear™.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.