CLEANING

1. Disengage power cord before performing any

cleaning.

2. The interior bottom, sides, shelves and trays of

the unit can be cleaned using a suitable cleaning
and disinfecting solution. Allow to dry thoroughly
before restarting unit.


3. The exterior can be cleaned with a damp cloth or

sponge.

4. Be sure to properly dispose of all cleaning materi-

als, especially any radioactive items.

SERVICE: RETURN AUTHORIZATION IS
REQUIRED.

Should service be required, contact your salesperson
or call Boekel Scientific Customer Service.

NOTE: Before shipping this unit for service, the

unit must

Size:

be thoroughly cleaned and decon-

taminated of any radioactivity. Shipment of
unmarked radioactive units is illegal. All in-
coming units will be tested for radioactivity.
Disposal and/or decontamination costs will
be charged for any radioactive unit returned
for repair.

SPECIFICATIONS

Exterior:

14.5”D (37 cm) x 16”H (41 cm) x

14W (36 cm)
Chamber:

11.5”D (29 cm) x 10.5”H (27 cm) x

11.5W (29 cm)
Tray:

8”D (20 cm) x 0.5”H (1.3 cm) x

8.5W (22 cm)

Shelf Separation: 2.5” (6.35 cm)
Shelf Area: 9”D (23 cm) x 9.5”W (24 cm)
Temperature Range: Ambient plus 5°C to 70°C Opera-

tion above 70°C not recommended.

Shaking Speed: 8 to 25 strokes per minute, typical
Weight: 24 lbs (11 kg)
Power: Model 136400, 136400S 115 VAC, 230 watts

Model 136400-2

230 VAC, 230 watts















TYPICAL CHAMBER TEMPERATURE PERFORMANCE

Condition

1

Chamber Stability

2

Chamber Uniformity

3

37°C without trays

37°C ±0.2°C

±0.5°C

37°C with trays

37°C ±0.5°C

±1.2°C

65°C without trays

65°C ±1.0°C

±1.3°C

65°C with trays

65°C ±1.0°C

±2.3°C

65°C in sealed bag or box on shelf – Uniformity of 65°C ±0.7°C


1 – With trays refers to brown plastic drip trays supplied with unit. When placed on shelves, trays can create some

temperature stratification.

2 – Chamber average of 5 test points within the shelf boundaries of the unit during operation.