1-4 Battery Condition and Operation

Below is a list of the battery operation according to the combination of the power switch and AC

adapter conditions. For details, read each section on the following pages.

Operation

Power
switch

AC

adapter

Description

Charging the battery

OFF

Connected

The CHARGER LED indicator blinks and
charging starts. After about 10 hours,
charging stops automatically and the
CHARGER LED indicator goes out.

Charge the battery after it is completely
discharged.

During the charging operation, the
battery can not be used.

Using the balance powered
by the battery

ON

Not
connected

The battery supplies power to the
balance.

The battery will supply power
continuously for 8 hours under normal
conditions.

None OFF

Not
connected

The battery is not used.

Using the balance powered
by the AC adapter

ON Connected

The AC adapter supplies power to the
balance.

The AC adapter does not charge the
battery.